Remote Commands
R&S
®
FPL1000
675
User Manual 1178.3370.02 ─ 03
Suffix:
<p>
.
1...4
Power sensor index
Usage:
Query only
[SENSe:]PMETer<p>:DCYCle[:STATe]
<State>
This command turns the duty cycle correction on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
PMET2:DCYC:STAT ON
Manual operation:
See
[SENSe:]PMETer<p>:DCYCle:VALue
<Percentage>
This command defines the duty cycle for the correction of pulse signals.
The power sensor uses the duty cycle in combination with the mean power to calculate
the power of the pulse.
Suffix:
<p>
.
1...4
Power sensor
Parameters:
<Percentage>
Range:
0.001 to 99.999
*RST:
99.999
Default unit: %
Example:
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
Manual operation:
See
[SENSe:]PMETer<p>:FREQuency
<Frequency>
This command defines the frequency of the power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Setting Basic Spectrum RF Measurement Parameters