Remote Commands
R&S
®
FPL1000
627
User Manual 1178.3370.02 ─ 03
Parameters:
<FilterType>
CFILter
NORMal
P5
RRC
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Return values:
<PulsePower>
List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Usage:
Query only
[SENSe:]MPOWer[:SEQuence]
<Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <#OfPulses>
This command configures and initiates the pulse power measurement.
The R&S
FPL1000 caches all measurement parameters that you can set with this
command. If you use the command repeatedly, the R&S
FPL1000 only changes those
settings that you have actually changed before initiating the measurement. Thus, mea-
surement times are kept as low as possible.
If you synchronize the measurement with
*OPC
, the R&S
FPL1000 produces a service
request when all frequencies have been measured and the number of individual mea-
surements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see
Parameters:
<Frequency>
Defines the pulse frequency.
Range:
0 to Fmax
Default unit: Hz
<RBW>
Defines the resolution bandwidth.
<MeasTime>
Defines the measurement time.
Range:
1 µs to 30 s
<TriggerSource>
Selects a trigger source.
Configuring and Performing Measurements