Remote Commands
R&S
®
FPL1000
570
User Manual 1178.3370.02 ─ 03
<range>
1..30
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
ESP:RANG2:INP:ATT:AUTO OFF
Deactivates the RF attenuation auto mode for range 2.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a SEM range on and off.
In case of high speed measurements, the state of the preamplifier has to be identical
for all ranges.
Suffix:
<sb>
.
1 | 2 | 3
<range>
1..30
Selects the measurement range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>:INSert
<Mode>
This command inserts a new SEM range and updates the range numbers accordingly.
Suffix:
<sb>
.
1 | 2 | 3
<range>
1..30
Selects the SEM range.
Parameters:
<Mode>
AFTer
Inserts a range after the selected range.
BEFore
Inserts a range before the selected range.
Manual operation:
See
" Insert before Range / Insert after Range "
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:ABSolute:STARt
<Level>
This command defines an absolute limit for a SEM range.
Configuring and Performing Measurements