Remote Commands
R&S
®
FPL1000
569
User Manual 1178.3370.02 ─ 03
●
smaller than the span the SEM sweep list covers, the R&S
FPL1000 will not mea-
sure the ranges that are outside the span - results may be invalid.
●
greater than the span the SEM sweep list covers, the R&S
FPL1000 will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
For more information see
Suffix:
<sb>
.
1 | 2 | 3
<range>
1...30
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>:INPut:ATTenuation
<Attenuation>
This command defines the input attenuation for a SEM range.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
.
1 | 2 | 3
<range>
1...20
Selects the measurement range.
Parameters:
<Attenuation>
Numeric value.
Refer to the data sheet for the attenuation range.
*RST:
10 dB
Default unit: dB
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>:INPut:ATTenuation:AUTO
<State>
This command turns automatic selection of the input attenuation for a SEM range on
and off.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
.
1 | 2 | 3
Configuring and Performing Measurements