Remote Commands
R&S
®
FPL1000
552
User Manual 1178.3370.02 ─ 03
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
0 dB to 100 dB
*RST:
0 dB
Default unit: dB
<UpperLimit>
The limit of the upper adjacent channel.
Range:
0 dB to 100 dB
*RST:
0 dB
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<k>:ACPower:ACHannel:RESult?
This command queries the state of the limit check for the adjacent channels in an
ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>, <k>
.
irrelevant
Return values:
<LowerACH>,
<UpperACH>
text value
The command returns two results. The first is the result for the
lower, the second for the upper adjacent channel.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage:
Query only
Manual operation:
See
CALCulate<n>:LIMit<k>:ACPower:ACHannel[:RELative]:STATe
<State>,<State>
This command turns the relative limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<k>:ACPower[:STATe]
.
Suffix:
<n>, <k>
.
irrelevant
Configuring and Performing Measurements