![Rohde & Schwarz FPL1000 Series User Manual Download Page 366](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826366.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
366
User Manual 1178.3370.02 ─ 03
With the R&S
FPL1000, the power sensors can be connected to the "Power Sensor"
interface directly, and no further cables are required. They can then be configured as
an external power sensor trigger.
Figure 7-35: Connecting a power sensor using the POWER SENSOR interface
The R&S
FPL1000 receives an external trigger signal when the defined trigger level is
measured by the power sensor. Power measurement results are provided as usual.
The "Gate Mode"
Level
is not supported for R&S power sensors. The signal sent by
these sensors merely reflects the instant the level is first exceeded, rather than a time
period. However, only time periods can be used for gating in level mode. Thus, the trig-
ger impulse from the sensors is not long enough for a fully gated measurement; the
measurement cannot be completed. For details on gating see
For details see
"How to Configure a Power Sensor as an External (PSE) Trigger"
7.3.2.2
Power Sensor Settings
Access
: "Overview" > "Input" > "Power Sensor" tab
Each sensor is configured on a separate tab.
Receiving Data Input and Providing Data Output