The Spectrum Application (RF Measurements)
R&S
®
FPL1000
290
User Manual 1178.3370.02 ─ 03
You can define in which detail the data is displayed in the result summary in the "List
Evaluation" settings (see
"List Evaluation (Results Configuration)"
default, one peak per range is displayed. However, you can change the settings to dis-
play only peaks that exceed a threshold ( "Margin" ).
Detected peaks are not only listed in the Result Summary, they are also indicated by
blue squares in the diagram (optionally, see
tings).
Figure 7-23: Detected peak display in SEM measurement
Furthermore, you can export the results of the result summary to a file which can be
exported to another application for further analysis.
Results for SEM with multiple sub blocks
In the Spectrum application only, spectrum emissions can be measured for multiple
sub blocks of channels (see
"SEM with Multiple Sub Blocks ("Multi-SEM")"
on page 297 for details). Up to three sub blocks (with two gaps) can be defined. For
each sub block and each gap, the results described above are provided individually in
the result summary.
Measurements and Results