![Rohde & Schwarz FPL1000 Series User Manual Download Page 19](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826019.webp)
Getting Started
R&S
®
FPL1000
19
User Manual 1178.3370.02 ─ 03
4.1.1.8
Switching the Instrument On and Off
Switching on the instrument
1. Press the AC power switch on the rear panel to position "I".
2. Press the POWER key on the front panel of the instrument.
The instrument operates on DC or AC power, whichever is supplied. After booting,
the instrument is ready for operation. The green POWER LED indicates this.
A steady orange LED indicates the instrument is in standby mode.
Warm-up time for OCXO (option R&S
FPL1-B4)
When the instrument is switched on, the OCXO requires an extended warm-up time
(see data sheet).
Switching off the instrument
1. Press the POWER key on the front panel.
The R&S
FPL1000 switches to standby mode (not in battery operation).
2. For AC power operation, set the AC power switch on the rear panel to position "O".
For DC power operation, disconnect the DC power supply cable.
The R&S
FPL1000 changes into off mode.
Risk of losing data
If you switch off the running instrument using the rear panel switch or by disconnecting
the power cord, and no (charged) battery is available, the instrument loses its current
settings. Furthermore, program data may be lost.
Press the Power key first to shut down the application properly.
4.1.1.9
Performing a Self-Alignment
During instrument start, the installed hardware is checked against the current firmware
version to ensure the hardware is supported. If not, an error message is displayed
( "Wrong Firmware Version" ) and you are asked to update the firmware. Until the firm-
ware version is updated, self-alignment fails.
(For details refer to the R&S
FPL1000 User Manual).
When strong temperature changes occur, you may have to perform a self-alignment to
align the data to a reference source.