Preset
R&S
®
FPC
37
User Manual 1178.4130.02 ─ 07
8
Preset
Access
: "Preset"
Access
(preset configuration): "Setup" > "User Preference"
Access
(factory reset): "Setup" > "Instrument Setup"
Remote commands required to preset the system:
●
Before you prepare a measurement, it is recommended to preset the R&S
FPC. During
a preset, the R&S
FPC resets all settings to their default state. Restoring the default
configuration has the advantage that old settings do not affect measurements.
The default setup depends on the application you are using.
Presetting the system
1. Press the "Preset" key to restore the default instrument configuration.
You can also define your own default settings via a dataset. The R&S
FPC restores
the configuration stored in the dataset after you press the "Preset" key instead of
the factory defaults.
2. Press the "Setup" key to open the setup menu.
3. Select the "User Preference" menu item to open the user preference dialog box.
4. Select the "Preset Mode" menu item to select the preset mode.
● "Default"
Restores the factory default configuration.
● "User"
Restores custom default values taken from a dataset. In that case, you have to
select a dataset that contains the default values.
5. Select the "Preset Dataset" menu item to select a dataset.
When you preset the R&S
FPC, it applies the configuration of the dataset.
For more information about the "Discard Calibration Data" menu item, see
forming a default calibration"
In addition to the preset, you can also reset the system to the condition it had when it
was delivered (factory reset).
Potential data loss
Note that when you reset the R&S
FPC to the factory state, all files you have saved on
the internal memory of the R&S
FPC will be deleted. Only the files that were included
in the delivery will be kept.
Resetting the system
1. Press the "Setup" key to open the setup menu.