Analog Demodulation
R&S
®
FPC
157
User Manual 1178.4130.02 ─ 07
3. Navigate to the dataset that contains the limits you want to use.
Make sure to select a dataset that is compatible to the current measurement: limits
for AM measurements have file extension
.amlim
, those for FM measurements
have the file extension
.fmlim
.
You can use the "Show Compatible" menu item to display only those files that you
can use for the current measurement mode.
4. Select one of the available files and confirm the selection with the "Load" menu
item.
The R&S
FPC shows the upper and lower limit values defined in the file in the cor-
responding columns of the result summary. It automatically checks if the results are
within the limits or not. Results within the limits are indicated by a "Pass" message
in a green font, results that violate the limits are indicated by a "Fail" message in a
red font.
Removing limits
1. Press the "Lines" key top open the limits menu.
2. Select the "Clear Limits" menu item.
The R&S
FPC removes the limit values from the result summary and no longer
evaluates them.
Turning an audio signal in case of limit violation on and off
1. Press the "Lines" key top open the limits menu.
2. Select the "Audio Beep" menu item.
The R&S
FPC plays an audio signal if any of the measured results fails the limit
check.