Remote Control Commands
R&S
®
EX-IQ-BOX
247
User Manual 1409.5570.32 ─ 05
:DIAGnostic:EBOX:LVDS:PLL?
Checks the PLL lock state during LVDS test. This function is protected and used for
service purposes.
Return values:
<pll>
ON | OFF
Example:
DIAG:EBOX:LVDS:PLL?
query the state of the PLL.
Usage:
Query only
:DIAGnostic:EBOX:LVDS:RCV?
The command checks the number of received bits.
Return values:
<received>
float
Example:
DIAG:EBOX:LVDS:RCV?
count the number of received bits.
Usage:
Query only
:DIAGnostic:EBOX:LVDS:RX|TX:SOURce
<source>
Selects LVDS test source. This function is protected and used for service purposes.
Parameters:
<source>
PRBS | COUNTER
Example:
DIAG:EBOX:LVDS:TX:SOUR PRBS
test the TX channel with the PRBS signal pattern.
:DIAGnostic:EBOX:LVDS:RX|TX:STATe
<state>
Starts the LVDS test. This function is protected and used for service purposes.
Parameters:
<state>
0 | 1 | OFF | ON
*RST:
OFF
Example:
DIAG:EBOX:LVDS:TX:STAT ON
start the LVDS test in the TX channel.
:DIAGnostic:EBOX:POINt
<point>
Selects the diagnostic points. This function is protected and used for service purposes.
R&S EX-IQ-BOX - Remote Control Commands