![Rohde & Schwarz EX-IQ-BOX User Manual Download Page 128](http://html1.mh-extra.com/html/rohde-and-schwarz/ex-iq-box/ex-iq-box_user-manual_1477982128.webp)
Protocol Settings
R&S
®
EX-IQ-BOX
113
User Manual 1409.5570.32 ─ 05
z 2´s Complement
Binary Offset
-4 1
1
0
0
0
1
0
0
-3 1
1
0
1
0
1
0
1
-2 1
1
1
0
0
1
1
0
-1 1
1
1
1
0
1
1
1
0 0
0
0
0
1
0
0
0
1 0
0
0
1
1
0
0
1
2 0
0
1
0
1
0
1
0
3 0
0
1
1
1
0
1
1
4 0
1
0
0
1
1
0
0
5 0
1
0
1
1
1
0
1
6 0
1
1
0
1
1
1
0
7 0
1
1
1
1
1
1
1
"2´s Comple-
ment"
The value of the transmitted data is formatted in two´s-complement.
The most significant bit has a value of -2
n-1
, the bits of lesser signifi-
cance follow as:
+2
n-2
...
+2
0
"Binary Offset"
The value of the transmitted data is formatted in binary offset.
A binary offset of -2
n-1
is added such that the final values are always
positive.
Example:
n = 4
→
-8
≤
z < 8
Remote command:
Strobe Polarity
Determine the polarity of the strobe marker signal. During a serial data transmission every
data sample is marked by the strobe marker signal.
The strobe marker is output at the UI_GP5 of the user interface.
Note:
This parameter applies to serial data transmission.
"Clock Phase" or "Clock Skew" settings
do not affect
the strobe marker signal.
User Defined