
Applications
R&S
®
ESW
92
User Manual 1177.6298.02 ─ 08
Example: Sequencer procedure
Assume the following active channel definition:
Tab name
Application
Sweep mode
Sweep count
Spectrum
Spectrum
Cont. Sweep
5
Spectrum 2
Spectrum
Single Sweep
6
Spectrum 3
Spectrum
Cont. Sweep
2
IQ Analyzer
IQ Analyzer
Single Sweep
7
For
Single Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer
For
Continuous Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
...
For
Channel-defined Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 2 x Spectrum 3,
5x Spectrum, 2 x Spectrum 3,
...
RUN SINGLE/RUN CONT and Single Sweep/Sweep Continuous keys
While the Sequencer is active, the [RUN SINGLE] and [RUN CONT] keys control the
Sequencer, not individual sweeps. [RUN SINGLE] starts the Sequencer in single
mode, while [RUN CONT] starts the Sequencer in continuous mode.
Running a Sequence of Measurements