Remote Control
R&S
®
ESRP
705
User Manual 1175.7445.02 ─ 08
The command is available only for multicarrier channel and adjacent-channel power
measurements with span > 0 (
CALCulate<n>:MARKer<m>:FUNCtion:POWer:
Parameters:
<Channel>
MINimum | MAXimum | LHIGhest
MINimum
Transmission channel with the lowest power
MAXimum
Transmission channel with the highest power
LHIGhest
Lowermost transmission channel for the lower adjacent chan-
nels, uppermost transmission channel for the upper adjacent
channels
Example:
POW:ACH:REF:TXCH:AUTO MAX
The transmission channel with the highest power is used as a
reference channel.
Manual operation:
See
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:MANual
<ChannelNumber>
This command selects a transmission channel to be used as a reference channel in
relative adjacent-channel power measurements.
The command is available only for multicarrier channel and adjacent-channel power
measurements with span > 0 (
CALCulate<n>:MARKer<m>:FUNCtion:POWer:
Parameters:
<ChannelNumber>
1 to 18
*RST:
1
Example:
POW:ACH:REF:TXCH:MAN 3
Transmission channel 3 is used as a reference channel.
Manual operation:
See
Checking Limits
The following commands configure limit checks for channel power and ACLR measure-
ments.
CALCulate<n>:LIMit<k>:ACPower:ACHannel:ABSolute
..................................................... 706
CALCulate<n>:LIMit<k>:ACPower:ACHannel:ABSolute:STATe
...........................................706
CALCulate<n>:LIMit<k>:ACPower:ACHannel[:RELative]
....................................................707
CALCulate<n>:LIMit<k>:ACPower:ACHannel[:RELative]:STATe
......................................... 708
CALCulate<n>:LIMit<k>:ACPower:ACHannel:RESult
........................................................ 708
CALCulate<n>:LIMit<k>:ACPower:ALTernate<Channel>:ABSolute
..................................... 709
CALCulate<n>:LIMit<k>:ACPower:ALTernate<Channel>:ABSolute:STATe
........................... 710
CALCulate<n>:LIMit<k>:ACPower:ALTernate<channel>[:RELative]
..................................... 711
Remote Commands in Spectrum Analyzer Mode