Remote Control
R&S
®
ESRP
678
User Manual 1175.7445.02 ─ 08
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:DOMain
<SpanSetting>
This command selects the domain of the limit line.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Parameters:
<SpanSetting>
FREQuency | TIME
*RST:
FREQuency
Example:
CALC:LIM2:CONT:DOM TIME
Defines zero span for the x-axis of limit line 2.
CALCulate<n>:LIMit<k>:CONTrol:MODE
<Mode>
This command selects the horizontal limit line scaling.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST:
ABSolute
Example:
CALC:LIM2:CONT:MODE REL
Defines the x-axis of limit line 2 as relatively scaled.
CALCulate<n>:LIMit<k>:CONTrol:OFFSet
<Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Remote Commands in Receiver Mode