Remote Control
R&S
®
ESRP
667
User Manual 1175.7445.02 ─ 08
Manual operation:
See
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT
This command positions the delta marker to the next smaller trace maximum. The cor-
responding delta marker is activated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
CALC:DELT2:MAX:NEXT
Sets delta marker 2 to the next smaller maximum value.
Manual operation:
See
See
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK]
This command positions the delta marker to the current trace maximum. If necessary,
the corresponding delta marker is activated first.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
CALC:DELT3:MAX
Sets delta marker 3 to the maximum value of the associated
trace.
Manual operation:
See
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt
This command positions the delta marker to the next smaller trace maximum on the
right of the current value (i.e. ascending X values). The corresponding delta marker is
activated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
CALC:DELT:MAX:RIGH
Sets delta marker 1 to the next smaller maximum value to the
right of the current value.
Remote Commands in Receiver Mode