DISPlay[:WINDow<n>]:TRACe<t>:X:SPACing
<Scale>
This command selects the scale of the frequency axis.
Parameters:
<Scale>
LINear
Linear scale of the frequency axis.
LOGarithmic
Logarithmic scale of the frequency axis.
*RST:
LOGarithmic
Example:
DISP:TRAC:X:SPAC LIN
Selects linear scale for the frequency axis.
Manual operation:
See
See
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the receiver frequency for measurements in the frequency or
time domain.
Parameters:
<Frequency>
0 to fmax
Range:
0 Hz to fmax
*RST:
fmax /2
Example:
FREQ:CENT 100MHz
Defines a receiver frequency of 100 MHz.
Manual operation:
See
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
Parameters:
<StepSize>
Range:
1 to fmax
*RST:
0.1 x <span value>
Default unit: Hz
Example:
FREQ:CENT:STEP 120 MHz
Manual operation:
See
See