System Configuration
R&S
®
ESRP
489
User Manual 1175.7445.02 ─ 08
Remote command:
GPIB Terminator LFEOI/EOI ← GPIB ← General Setup
Changes the GPIB receive terminator.
According to the standard, the terminator in ASCII is <LF> and/or <EOI>. For binary
data transfers (e.g. trace data) from the control computer to the instrument, the binary
code used for <LF> might be included in the binary data block. Therefore, it should not
be interpreted as a terminator in this particular case. This can be avoided by changing
the receive terminator to EOI.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
*IDN Format Leg./New ← GPIB ← General Setup
Defines the response format to the
*IDN?
remote command (see
on page 884). This function is intended for re-use of existing control programs
together with the R&S
ESRP.
"Leg"
Legacy format, compatible to the R&S FSP/FSU/FSQ family
e.g.
Rohde&Schwarz,FSV-7,100005/007,1.61
"New"
R&S
ESRP format
e.g.
Rohde&Schwarz,FSV-7,1307.9002K07/100005,1.61
Remote command:
I/O Logging (On/Off) ← GPIB ← General Setup
Activates or deactivates the SCPI log function. All remote control commands received
by the R&S
ESRP are recorded in the following log file:
C:\R_S\Instr\scpilogging\ScpiLog.txt
Logging the commands can be extremely useful for debug purposes, e.g. to find mis-
spelled keywords in control programs.
Remote command:
Time+Date ← General Setup
Opens an edit dialog box to enter time and date for the internal real-time clock.
Chapter 3.2.4.2, "Setting the Date and Time"
Remote command:
Configure Monitor ← General Setup
Determines and displays the configuration of a connected external monitor, if available.
In the configuration dialog box, you can switch from the internal monitor (laptop icon) to
the external monitor (monitor icon), or both (double monitor icon). For external, the
R&S
ESRP display is disabled (turns dark). The screen content formerly displayed on
the R&S
ESRP is displayed on the external screen.
Instrument Setup and Interface Configuration – SETUP Key