Spectrum Measurements
R&S
®
ESRP
323
User Manual 1175.7445.02 ─ 08
Ref Level
Opens an edit dialog box to enter the reference level in the current unit (dBm, dBµV,
etc).
The reference level is the maximum value the AD converter can handle without distor-
tion of the measured value. Signal levels above this value will not be measured cor-
rectly, which is indicated by the "IFOVL" status display.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
Range
Opens a submenu to define the display range of the level axis.
●
●
●
●
●
●
●
●
Range Log 100 dB ← Range
Sets the level display range to 100 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 100DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log 50 dB ← Range
Sets the level display range to 50 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 50DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log 10 dB ← Range
Sets the level display range to 10 dB.
Configuration