Spectrum Measurements
R&S
®
ESRP
308
User Manual 1175.7445.02 ─ 08
Deltamarker 3, on the other hand, can be moved for fine adjustment irrespective of
marker 2.
Marker 1 can also be moved manually for re-adjustment without affecting the position
of the deltamarkers.
For general information on measuring the AM modulation depth see
"Measuring the AM Modulation Depth"
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult?
Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker Norm/Delta ← AM Mod
Depth
The "Marker X" softkey activates the corresponding marker and opens an edit dialog
box to enter a value for the marker to be set to. Pressing the softkey again deactivates
the selected marker.
If a marker value is changed using the rotary knob, the step size is defined via the
Marker 1 is always the reference marker for relative measurements. If activated, mark-
ers 2 to 16 are delta markers that refer to marker 1. These markers can be converted
into markers with absolute value display using the "Marker Norm/Delta" softkey. If
marker 1 is the active marker, pressing the "Marker Norm/Delta" softkey switches on
an additional delta marker.
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
CALCulate<n>:DELTamarker<m>:X:RELative?
Search Signals ← AM Mod Depth
Activates all markers.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:SEARchsignal ONCE
6.1.1.12
Measuring Harmonic Distortion
The harmonics and their distortion can be measured using the
With this measurement it is possible to measure the harmonics e.g. from a VCO easily.
In addition the THD (total harmonic distortion) is calculated in % and dB.
With span > 0 Hz, an automatic search for the first harmonic is carried out within the
set frequency range. Also the level is adjusted. In zero span, the center frequency is
unchanged.
Measurements