Receiver Mode
R&S
®
ESRP
180
User Manual 1175.7445.02 ─ 08
Marker Track
Turns marker frequency tracking on and off.
If on, the R&S
ESRP changes the center frequency to the marker frequency when you
change the marker position.
Remote command:
CALCulate<n>:MARKer<m>:COUPled[:STATe]
Marker to Stepsize
Matches the
to the marker frequency.
Example:
If the marker is positioned on a frequency of 1.25
MHz, the frequency stepsize is
1.25
MHz.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CSTep
Settings Coupled
Couples or decouples the receiver settings to the scan range settings when you use
"Marker Tracking".
If on, the R&S
ESRP changes the receiver settings according to the scan range the
marker frequency is currently in.
Remote command:
CALCulate<n>:MARKer<m>:SCOupled[:STATe]
5.4.2.2
Positioning Markers
The MKR
➙
key is used for search functions of measurement markers, assignment of
the marker frequency as center frequency, restriction of the search area and character-
ization of maxima and minima. For details on markers in general, see
To open the Marker To menu
●
Press the MKR -> key.
The "Marker To" menu is displayed. If no marker is active, marker 1 will be activa-
ted and a peak search on the trace carried out. Otherwise, the edit dialog box for
the last activated marker is opened and the current frequency/time value is dis-
played.
Further information
●
"Effect of Different Peak Excursion Settings (Example)"
Tasks
●
●
Common Analysis Functions