R&S ENV432
Introduction
User Manual 1326.6492.04 - 03
E-9
2.1.3 Schuko Socket (EUT 1)
Connector for disturbance voltage measurements on single-phase EUTs with an
alternating voltage up to 240 V and direct voltages up to 350 V. When using this
connector, the maximum constant current is 16 A.
2.1.4 CEE Socket (EUT 2)
Connector for disturbance voltage measurements on three-phase EUTs with a star-
type alternating voltage up to 240 V (corresponds to a delta-type alternating voltage of
415 V) and direct voltages up to 350 V. The maximum constant current per phase is
32 A.
2.1.5 Fan
Pushing the fan button activates the ventilation manually to prevent any damage due to
overheating. The V-network ventilation is automatically set to full speed at an internal
housing temperature of approx. 50 °C and the fan LED turns orange. If the coil
temperature is > 100 °C, an acoustic alarm sounds and the fan LED turns red.
2.1.6 Limiter
Pushing the limiter button switches off the pulse limiter.
2.1.7 Line Selection
Rotary switch for selecting the test path (L1, L2, L3 and N) of the V-network to the test
receiver. Depending on the test path selection the corresponding green line LED light
up.
2.1.8 RF Output to Test Receiver
Output N 50
, female, for connecting the test receiver.
2.1.9 Protective Earth Conductor Terminal (
)
M8 threaded bolt for connecting the protective earth conductor.