The R&S ENV432 four-line V-network is used to measure disturbance voltages on
mains-dependent loads.
Its main functions are as follows:
● Supply the equipment under test (EUT) with AC supply voltage or DC supply
voltage
● Provide a standardized load impedance
● Deliver the unsymmetric disturbance voltage generated by the EUT to the test
receiver in a defined manner
● Decouple the test circuit from mains disturbances
The R&S ENV432 V-network is built with air-core inductances (50
μH and 250 μH) and
conforms to the requirements specified in CISPR 16-1-2.
Additional equipment:
● Artificial hand
● 10 dB attenuator in decoupling branch
● Pulse voltage limiter at measurement output
● Remote control interface
● Separate power supply for control electronics and remote control interface
The V-network can be operated manually or under remote control via a TTL interface.