Basic Instrument Functions
R&S
®
CMW500
86
User Manual 1173.9463.02 ─ 02
●
The instrument setup
SCPI command:
etc.
Reset
Sets the instrument parameters to values suitable for remote operation. In particular, the
"Reset" state comprises the following settings:
●
All measurements are performed in Single-Shot mode ("Repetition: Single-Shot").
●
The R&S CMW500 uses short statistics cycles (for benchmarks).
●
In multi evaluation measurements, some time-consuming evaluations are skipped to
gain measurement speed.
●
The preferred modes and features are the same in all software versions. As far as
older features are concerned, the Reset state never changes after a software
upgrade.
Exceptions:
The following R&S CMW500 settings are not affected by "Reset":
●
Address information assigned to the instrument (e.g. the IP address)
●
The instrument setup
●
The contents of the status registers
SCPI command:
etc.
Factory Default
Sets the instrument to its factory default state. This state comprises the following settings:
●
settings: The instrument is optimized for local/manual operation.
●
The address information assigned to the instrument (e.g. the IP address or USB
product ID/vendor ID/device ID).
●
The instrument setup: One sub-instrument.
Use the
to restore the factory default settings.
SCPI command:
n/a
5.2.2 Save/Recall Dialog
The "Save/Recall" dialog stores the current instrument setup and recalls (and re-acti-
vates) setups.
To open the dialog press SAVE.
Utility Dialogs