Command Reference
R&S
®
CMW500
261
User Manual 1173.9463.02 ─ 02
TRACe:REMote:MODE:FILE<inst>:FILTer
<Input>, <Output>, <Error>, <Trigger>,
<DeviceClear>, <StatusRegister>, <Connection>, <RemoteLocal>
Specifies a filter for tracing of the specified sub-instrument. The filter defines which mes-
sage types and events are traced into a file.
The default setting is ON,ON,ON,OFF,OFF,OFF,OFF,OFF.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<Input>
ON | OFF
Trace incoming commands
<Output>
ON | OFF
Trace outgoing responses to queries
<Error>
ON | OFF
Trace SCPI error messages
<Trigger>
ON | OFF
Trace trigger events
<DeviceClear>
ON | OFF
Trace device clear messages
<StatusRegister>
ON | OFF
Trace status register changes
<Connection>
ON | OFF
Trace remote connection changes
<RemoteLocal>
ON | OFF
Trace transitions between local and remote operation mode
Example:
TRACe:REMote:MODE:FILE1:FILTer
ON,ON,ON,OFF,OFF,OFF,OFF,ON
Trace incoming commands, outgoing responses to queries, error
messages and remote-local transitions for sub-instrument 1.
Firmware/Software:
V2.0.10
TRACe:REMote:MODE:FILE<inst>:FORMat
<Format>
Specifies the format of the target file for tracing of the remote interface for the specified
sub-instrument. The trace can be stored as ASCII file or as XML file.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Instrument-Control Commands