![Rohde & Schwarz CMW500 Series User Manual Download Page 140](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718140.webp)
Remote Control
R&S
®
CMW500
140
User Manual 1173.9463.02 ─ 02
The relationship between measurement states and measurement control commands is
shown in the following diagram:
Measurement control commands are of the following type (see also
SCPI subsys-
tem
<Application>,
e.g.
Measurement instance
Context
INITiate
ABORt
STOP
:WIMax |
:GPRF
:MEASurement<i>
:POWer
:EPSensor
:MEValuation
...
(depending on <Application>)
Example
:
INITiate:GPRF:MEASurement:POWer
INITiate:<Application>:MEASurement<i>
Starts the measurement in the repetition mode set via
CONFigure:<Application>:MEASurement<i>:<Context>:REPetition
(sin-
). The command resets the counters for
the evaluation period and statistics cycle to zero and starts a timer. If the measurement
has not completed the first measurement cycle when the timer expires, the measurement
is stopped. The timeout value can be configured individually for each measurement.
INITiate...
can be used irrespective of the current measurement state. An initiated
measurement reserves all necessary system resources and enters the "RUN" state. If
Control of the Instrument