Command Reference
R&S
®
CMW-KM750/-KM751
91
User Manual 1173.9670.02 ─ 05
<EnableMagError>
OFF | ON
Disable or enable measurement of magnitude error
*RST:
ON
<EnablePhaseError> OFF | ON
Disable or enable measurement of phase error
*RST:
ON
<EnableIQ>
OFF | ON
Disable or enable measurement of I/Q origin offset and imbalance
*RST:
OFF
<EnableModScalar> OFF | ON
Disable or enable measurement of the other modulation results
*RST:
OFF
<EnablePMonitor>
OFF | ON
Disable or enable the power monitor
*RST:
OFF
Example:
See
Firmware/Software:
V3.0.10
V3.0.30: maximum number of segments increased to 500
Options:
R&S CMW-KM012
CONFigure:TDSCdma:MEAS<i>:MEValuation:LIST:SEGMent<Segno>:SPECtrum
<SpecStatistics>, <EnableACLR>, <EnableEMask>, <EnableOBW>
Defines the statistical length for the
AVERage
and
MAXimum
calculation and enables the
calculation of the different spectrum results in segment <Segno>.
Suffix:
<Segno>
.
1..500
Parameters:
<SpecStatistics>
Statistical length in slots. The maximum possible value depends
on the number of subframes in the segment and on the number of
active uplink slots per subframe, see
on page 88.
Range:
1 to 1000
*RST:
10
<EnableACLR>
OFF | ON
OFF:
Disable ACLR measurement
ON:
Enable ACLR measurement
*RST:
ON
Multi Evaluation Measurement Commands