CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
846
User Manual 1173.9511.02 ─ 12
Firmware/Software:
V2.0.10
V2.1.20: Support for state "Aliased"
CALCulate:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:CPO:CURRent?
CALCulate:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:CPO:AVERage?
CALCulate:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:CPO:MAXimum?
FETCh:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:CPO:CURRent?
FETCh:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:CPO:AVERage?
FETCh:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:CPO:MAXimum?
Returns Channel Phase Offset (CPO) results for the segment <no> in list mode (see
CONFigure:CDMA:MEAS<i>:MEValuation:LIST
). To define the statistical length
for
AVERage
and
MAXimum
calculation and enable the calculation of the results use the
CONFigure:CDMA:MEAS<i>:MEValuation:LIST:SEGMent<no>:
.
The values described below are returned by
FETCh
commands.
CALCulate
com-
mands return limit check results instead, one value for each result listed below.
The number to the left of each result parameter is provided for easy identification of the
parameter position within the result array.
Suffix:
<no>
.
1..200
The segment number must not exceed the total number of seg-
ments measured.
Return values:
<1_Reliability>
. In list mode, a zero reliability indicator indi-
cates that the results in all measured segments are valid. A non-
zero value indicates that an error occurred in at least one of the
measured segments.
<2_SegReliability>
0 | 3 | 4 | 8
The segment reliability indicates whether one of the following
exceptions occurred in this segment:
0:
No error
3:
Signal overflow
4:
Signal low
8:
Synchronization error
If a combination of exceptions occurs, the most severe error is
indicated.
<3_RPICH>
<4_RDCCH>
<5_RCCCH>
<6_REACH>
<7_RFCH>
<8_RSCH0w02E04>
<9_RSCH0w01E02>
<10_RSCH1w06E08>
<11_RSCH1w02E04>
Phase offset for the indicated channels in the in-phase signal
path (I-signal).
Range: –180 deg to +180 deg
Default unit: deg
Command Reference