CDMA2000 Signaling
R&S
®
CMW-KM8xx/-KS8xx
633
User Manual 1173.9511.02 ─ 12
Manual operation:
See
CONFigure:CDMA:SIGN<i>:RFPower:OUTPut?
Queries the total output power. The total output power includes the AWGN power level.
The allowed value:
Range (Output Power) = -130 dBm to -5 dBm (RFx Com) or -120
dBm to 3 dBm (RFx Out)
; please also notice the ranges quoted in the data sheet.
Return values:
<OutputPower>
Range:
-100 dBm to 100 dBm
Default unit: dBm
Example:
See
Specifying General Signaling Settings
Usage:
Query only
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:CDMA:SIGN<i>:RFPower:EPMode
<ExpPowerMode>
Configures the input path of the RF analyzer according to the expected output power of
the MS under test. The R&S
CMW assumes a 9 dB peak-to-average ratio (crest factor)
of the received CDMA2000 signal and allows for an additional reserve.
See also:
Parameters:
<ExpPowerMode>
MANual | OLRule | MAX | MIN
MANual
: Assume that the MS transmits at the fixed "Manual
Expected Power" value and configure the R&S
CMW input path
accordingly.
OLRule
: Open Loop Rule: Assume that the MS transmits
according to the open loop power rule: The sum of the mean
input power at the MS receiver plus the mean output power at
the MS transmitter is maintained at a constant "power offset"
value: <input power> + <output power> = <power offset>. The
power offset depends on the band class; see 3GPP2 C.S0057-
D.
MAX
: Maximum: Assume that MS transmits at its maximum out-
put power (RMS value
≤
+23 dBm).
MIN
: Minimum: Assume that MS transmits at its minimum output
power (RMS value
≤–
47 dBm).
*RST:
OLR
Example:
See
Specifying General Signaling Settings
Firmware/Software:
V1.0.15.0
Manual operation:
See
Command Reference