
CDMA2000 Signaling
R&S
®
CMW-KM8xx/-KS8xx
569
User Manual 1173.9511.02 ─ 12
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Single-shot is preferable if only a single measurement result is required under fixed
conditions, which is typical for remote-controlled measurements. Continuous mode is
suitable for monitoring the evolution of the measurement results in time and observe
how they depend on the measurement configuration, which is typically done in manual
control. The reset/preset values therefore differ from each other.
Remote command:
CONFigure:CDMA:SIGN<i>:RXQuality:FERFch:REPetition
CONFigure:CDMA:SIGN<i>:RXQuality:FERSch:REPetition
Stop Condition ← FER FCH / FER SCH0
Specifies the conditions for an early termination of the measurement:
●
None
: The measurement is performed according to its "Repetition" mode and "Max
Frames", irrespective of the limit check results.
●
Any Limit exceeded
: The measurement is stopped as soon as one of the limits is
exceeded, irrespective of the repetition mode set. If no limit failure occurs, it is per-
formed according to its "Repetition" mode and "Max Frames" settings. Use this
value for measurements that are essentially intended for checking limits, e.g. pro-
duction tests.
●
Min Confidence Level exceeded
: The measurement is stopped as soon as the
minimum confidence level exceeds the set limit. This indicates the probability that
the true FER is less than the set FER limit.
●
Max FER
: The measurement is stopped as soon as the number of frame errors
exceed the set limit .
Remote command:
CONFigure:CDMA:SIGN<i>:RXQuality:FERFch:SCONdition
CONFigure:CDMA:SIGN<i>:RXQuality:FERSch:SCONdition
Max Frames ← FER FCH / FER SCH0
The "Max Frames" sets the number of frames used to calculate the FER. Hence it
defines the length of a single shot FER measurement.
Remote command:
CONFigure:CDMA:SIGN<i>:RXQuality:FERFch:FRAMes
CONFigure:CDMA:SIGN<i>:RXQuality:FERSch:FRAMes
RLP > Window Size
Sets the active window size in an RLP mesurement.
Remote command:
CONFigure:CDMA:SIGN<i>:RXQuality:WINDowsize
Pilot Strength
Configures measurement control for measurements of pilot stregth.
Repetition ← Pilot Strength
Defines how often the measurement is repeated if it is not stopped explicitly or by a
failed limit check.
GUI Reference
135-1095-0799