1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
373
User Manual 1173.9511.02 ─ 12
Parameters:
<CMWSConnector> Selects the input connector of the R&S
CMWS
*RST:
R11
Example:
See
Firmware/Software:
V3.2.80
Options:
R&S CMW-KM012
CONFigure:EVDO:MEAS<i>:MEValuation:LIST:SEGMent<no>:SETup
<SegmentLength>, <Level>, <Frequency>, <RetriggerOption>, <PLSubtype>,
<HalfSlot>, <LongCodeMaskI>, <LongCodeMaskQ>[, <CMWSConnector>]
Defines the length of segment <no> and the analyzer settings. In general this com-
mand must be sent for all segments measured.
Suffix:
<no>
.
1..200
The segment number must not exceed the total number of seg-
ments measured.
Parameters:
<SegmentLength>
Number of measured slots in the segment.
The sum of the segment lengths must not exceed 8192 slots;
limit violation is indicated by a reliability indicator of 2 (Capture
Buffer Overflow).
Range:
1 to 1000
*RST:
1
<Level>
Expected nominal power in the segment. The range of the
expected nominal power can be calculated as follows:
Range (Expected Nominal Power) = Range (Input Power) +
External Attenuation - User Margin
The input power range is stated in the data sheet.
*RST:
0 dBm
<Frequency>
Center frequency of the RF analyzer.
Range:
100E+6 Hz to 6E+9 Hz
*RST:
833.49E+6 Hz
Default unit: Hz
Command Reference