Radar Echo Generation
R&S
®
AREG100A
51
User Manual 1178.7417.02 ─ 04
Figure 4-3: Analyzing Radar Signal
1. Connect an R&S FSW to the RX IF OUT connector of the R&S
AREG100A base
unit.
2. Set the center frequency of the R&S FSW appropriately. A radar signal at fre-
quency f_RF is output at an intermediate frequency f_IF at the RX IF out connector
according to the following equation:
f_IF = f_RF - f_LO
f_LO = 23.3 GHz for B124 options
f_LO = 75.3 GHz for B177 options
For B181 options, f_LO depends on the frequency setting.
● If the frequency setting for the R&S
AREG100A is below 78.5 GHz, f_LO =
75.3 GHz.
● If the frequency setting for the R&S
AREG100A is above 78.5 GHz, f_LO )
76.275 GHz.
3. The sending signal of the RUT can now be observed and evaluated.
4.2
GUI Elements
The R&S
AREG100A GUI consists of the following parts:
●
Frequency
●
Radar Objects
●
AREG Config
●
System Config
●
Power Sensors
GUI Elements