Arbitrary and Function Generator
R&S AM300
E-1007.9922.12
6-92
Operating manual, 02/2005
Setting the Signal Amplitude
Application
You can set the level of the
Pulse
waveform via the absolute value of the
amplitude V
p
.
Setting the
absolute value of
the amplitude V
p
1.
In the
submenu
press the
function key
.
An entry field containing the current setting is displayed. The default
setting is 100 mV.
2.
Enter a new value ( 5-55).
The permissible entry range for the absolute value is:
0 mV
≤
V
p
≤
10 V in 0.1 mV steps
Level display
The new setting is stored and then displayed in the diagram area. The
divisions on the level axis (grid lines) change in accordance with the
amplitude setting.
The currently applicable signal amplitude is displayed in the parameter
field below the channel name CH1 or CH2, e.g. for CH1 "Ampl: 5 Vp
Norm". The extension "Norm" indicates that the pulse is not inverted (
6-108).
Restricting the
value range
The entry range for the signal amplitude can be restricted by the
following parameters:
DC component DC OFFSET ( 6-81)
Limitation of output voltage LOWER/UPPER LIMIT
( 6-83)
Vp