Remote Control
R&S
®
NRT2
123
User Manual 1178.5550.02 ─ 04
Parameters:
<value>
Range:
0.0 to 1.0
*RST:
0.5
Default unit: -
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:RFRatio
<value>
Relative value for ratio of forward/reverse power.
Suffix:
<Measurement>
.
1...100
<DirectionalChannel> 1..2
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
50.0
Default unit: pct
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:RLOSs
<value>
Relative value for return loss.
Suffix:
<Measurement>
.
1...100
<DirectionalChannel> 1..2
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dB
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio:SWR
<value>
Relative value for SWR.
Suffix:
<Measurement>
.
1...100
<DirectionalChannel> 1..2
CALCulate