Measurement and display configuration
R&S
®
NRX
92
User Manual 1178.5566.02 ─ 09
Graphical display
Figure 6-7: Time gate, graphical display
In the upper pane, the measurement result is shown as a single trace measurement of
the primary sensor. The active gate is indicated as colored area. The color is matching
the color assigned to the gate. You can change the gate configuration directly by using
touch gestures. If you have a channel calculation function running that involves
2
power sensors, a legend in the upper right corner shows the assignment of curve to
measurement channel.
In the lower pane, the 4
gates are displayed with:
●
Assigned color
●
Measurement values:
–
"Av"
Average power within gate
–
"Pk"
Peak power within gate
–
"Cf"
Crest factor
●
Borders, see
.
●
Length
Δ
t
If you tap the lower pane, the
6.6.2
Time gate settings
Access: "Measurement Settings"
> "Measurement Type"
> "Time Gate"
The available settings depend on the selected result display.
Time gate