Remote control commands
R&S
®
NRX
315
User Manual 1178.5566.02 ─ 09
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
100.0
Default unit: pct
Manual operation:
See
CALCulate<Measurement>:STATistics[:SCALe]:X:MODE
<value>
Effective for statistics measurements.
Sets relative or absolute scaling for the x-axis.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
ABSolute | RELative
*RST:
ABSolute
Manual operation:
See
CALCulate<Measurement>:STATistics[:SCALe]:X:POINts
<value>
Effective for statistics measurements.
Sets the measurement result resolution. It specifies the number of pixels that are
assigned to the logarithmic level range for measured value output. The width of the
level range divided by N–1, where N is the number of pixels, must not be less than the
smallest level resolution.
●
CALCulate<Measurement>:STATistics[:SCALe]:X:RANGe
●
Smallest level resolution:
[SENSe<Sensor>:]STATistics:SCALe:X:
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
3 to 8191
*RST:
600
Manual operation:
See
Measurement settings and results