Remote control commands
R&S
®
NRX
245
User Manual 1178.5566.02 ─ 09
[SENSe<Sensor>:]POWer:REFerence
<ref>
Sets the reference value for the relative power indication.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<ref>
Range:
-290.0 to +110.0
*RST:
+30.0
Default unit: dBm
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
on page 58
See
13.6.6
Continuous average measurements
There are several ways to programm the R&S NRX for measurements. See
ter 13.4, "Making measurements"
Further information:
●
Chapter 6.2, "Continuous average"
13.6.6.1
Combining commands for continuous average measurements
See also
Chapter 13.4, "Making measurements"
Parameter list for continuous average measurements
Parameter
Description
<expected_value>
Optional. Value that is expected for the measurement.
<resolution>
Optional. Limit up to which the measurement result probably is free of
noise.
[SENSe<Sensor>:]AVERage:COUNt:AUTO:
<source_list>
Mandatory. Primary and secondary sensor. The number for each sensor
is preceded by the character
@
. The entire expression is enclosed in
parentheses.
Example: (
@
3),(
@
2)
Sensor
C is the primary sensor, and sensor
B is the secondary sensor.
Measurement settings and results