Remote control commands
R&S
®
NRX
225
User Manual 1178.5566.02 ─ 09
Parameters:
<value>
Manual operation:
See
See
"Forward Upper Limit, Reflection Upper Limit"
CALCulate<Measurement>:LIMit<DirectionalChannel>:UPPer[:DATA]:CCDF
<value>
Sets the upper limit for the complementary cumulative distribution function (CCDF).
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
100.0
Default unit: pct
Manual operation:
See
"Forward Upper Limit, Reflection Upper Limit"
CALCulate<Measurement>:LIMit<DirectionalChannel>:UPPer[:DATA]:RATio:
RCOefficient
<value>
Sets the upper limit for the reflection coefficient.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
5.0
Default unit: -
Manual operation:
See
See
"Forward Upper Limit, Reflection Upper Limit"
CALCulate<Measurement>:LIMit<DirectionalChannel>:UPPer[:DATA]:RATio:
RFRatio
<value>
Sets the upper limit for the ratio of forward/reverse power.
Measurement settings and results