Remote Control Basics
R&S
®
NRX
486
User Manual 1178.5566.02 ─ 08
Table 15-16: Used operation trigger extended status bits and their meaning
Bit no.
Short description
Bit is set if a measurement channel is waiting for trigger
event or for the extension summary.
0
Extension summary
Sum bit of the next lower status register.
1
Measurement m+1 wait for trigger
Channel
1
2
Measurement m+2 wait for trigger
Channel
2
3
Measurement m+3 wait for trigger
Channel
3
4
Measurement m+4 wait for trigger
Channel
4
5
Measurement m+5 wait for trigger
Channel
5
6
Measurement m+6 wait for trigger
Channel
6
7
Measurement m+7 wait for trigger
Channel
7
8
Measurement m+8 wait for trigger
Channel
8
15.4.8.6
Operation Sense Status Register
The
CONDition
register contains information whether a power sensor is being initial-
ized. The
EVENt
register contains information whether an initialization was started or
completed since the last query.
A power sensor is initialized if:
●
Supply voltage is switched on (power-up).
●
Sensor was connected.
●
.
Status Reporting System
www.allice.de
Allice Messtechnik GmbH