Remote control commands
R&S
®
RTM3000
769
User Manual 1335.9090.02 ─ 09
Parameters:
<State>
ON | OFF
*RST:
OFF
DIGital<m>:PROBe[:ENABle]?
Tests whether the digital probe is connected and recognized by the instrument.
Suffix:
<m>
.
0..15
For pod 1: a value between 0 and 7
For pod 2: a value between 8 and 15
Return values:
<ProbeEnable>
ON | OFF
Usage:
Query only
DIGital<m>:DISPlay
<State>
Enables and displays the indicated logic channel, or disables it.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameters:
<State>
ON | OFF
*RST:
OFF
DIGital<m>:TECHnology
<ThresholdMode>
Selects the threshold voltage for various types of integrated circuits and applies it to
the channel group to which the indicated logic channel belongs.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameters:
<ThresholdMode>
TTL | ECL | CMOS | MANual
TTL: 1.4 V
ECL: -1.3 V
CMOS: 2.5 V
MANual: Set a user-defined threshold value with
*RST:
MAN
Mixed signal option (option R&S
RTM-B1)