GUI Reference
R&S
®
ZNA
464
User Manual 1178.6462.02 ─ 12
In the converter configuration above, P1 and P2 drive the RF IN ports of converter
1 and 2, respectively, and the Converter LO port drives the LO IN ports of both con-
verters.
If a converter port is
not
, and a source flatness cali-
bration is made at both the RF IN and the converter port, then the flatness cal at
the converter port is enabled and the flatness cal at the RF IN port is disabled.
Select a "Ref. Receiver" or "Meas. Receiver" or "Source Flatness" symbol to start the
related calibration. See
"Power Cal Dialog (Ref. Receiver)"
"Power Cal Dialog (Source Flatness)"
A green checkmark indicates that the calibration data has been acquired successfully.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Apply
Is enabled when a new set of power calibration data has been acquired for either a
"Power" source or a "Meas. Receiver". The button applies all available source and/or
measurement receiver calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
Cal Softtool