GUI Reference
R&S
®
ZNA
459
User Manual 1178.6462.02 ─ 12
See also
Chapter 5.5.5.3, "Characterization of Calibration Units"
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Mixer/IMD/Harmonics Tab
Defines channel-specific settings for mixer and intermodulation (IMD) measurements.
Calibration Mixer Settings
Defines the properties of the calibration mixer and the related calibration. The table is
enabled, if the channel is configured for a mixer measurement.
IMD Cal Method>Fast Cal Interp. Meas
This checkbox is enabled, if the channel is configured for an intermodulation (IMD)
measurement.
To avoid interpolation of calibration data, which causes problems during the measure-
ment, it is preferable to increase the number of sweep points during the calibration.
However, the number of sweep points that are required to avoid interpolation can be
large, and hence the calibration can take some time to complete.
Cal Softtool