Command Reference
R&S
®
ZNA
1367
User Manual 1178.6462.02 ─ 12
Setting parameters:
<Sensor>
ASENsor | BSENsor
ASENsor
–
external power meter Pmtr 1.
BSENsor
–
external power meter Pmtr 2.
*RST:
ASENsor
Example:
SOUR:POW3:CORR:COLL BSEN
Perform a source power calibration for port 3 using power meter
no. 2.
SOUR:POW:CORR:PMET:ID 2
Select power meter no. 2.
SOUR:POW:CORR:ACQ PORT,3
Perform a source power calibration for port 3 using the previ-
ously selected power meter no. 2.
Usage:
Setting only
Manual operation:
See
TRACe:CLEar
<MemTrace>
Deletes one of the memory traces
Mem<n>[Trc<m>]
, where n = 1, ... 8.
Setting parameters:
<MemTrace>
MDATa1 | MDATa2 | MDATa3 | MDATa4 | MDATa5 | MDATa6 |
MDATa7 | MDATa8
Identifier for the memory trace; see
Range:
MDATA<n> where <n> = 1 to 8.
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1
).
TRAC:COPY "Mem_Pt20",CH1DATA
Copy the current state of the created trace to a memory trace
named "Mem_Pt20". The memory trace is not displayed.
DISP:WIND:TRAC2:FEED "MEM_PT20"
Display the created memory trace in the active diagram area
(diagram area no. 1).
Usage:
Setting only
TRACe[:DATA][:RESPonse][:ALL]?
<Response>
Returns the response values of the active data trace or memory trace (see
Note:
To read the response values of an arbitrary data or memory trace, use
. To read the response values of a trace acquired in single
OFF
.