Command Reference
R&S
®
ZNA
1337
User Manual 1178.6462.02 ─ 12
Parameters:
<FreqOffset>
Frequency offset.
Range:
0 Hz to 25 MHz
Increment: 1 Hz
*RST:
0 Hz
Default unit: Hz
Example:
FREQ:STAR 2GHz; STOP 3 GHz
Define a sweep range between 2 GHz and 3 GHz.
FREQ:OFFS:PWAV 10kHz; WAV 20kHz
Modify the receiver frequency settings.
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:WAVes
<FreqOffset>
Defines the constant receiver frequency offset for unprimed wave quantities a, b rela-
tive to the common receiver frequency.
Suffix:
<Ch>
.
Channel number.
<Pt>
Test port number of the analyzer.
Parameters:
<FreqOffset>
Frequency offset.
Range:
0 Hz to 25 MHz
Increment: 1 Hz
*RST:
0 Hz
Default unit: Hz
Example:
See
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes
8.3.20.3
Internal 2nd LO (R&S
ZNA-B5)
[SENSe<Ch>:]MSMode
<LO Sequence Mode>
Suffix:
<Ch>
.
Parameters:
<LO Sequence
Mode>
ALTernated | CHOPped
[SENSe<Ch>:]SLAMode
<Port to LO Assignment Mode>
Defines the LO usage for channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Port to LO
Assignment Mode>
OPTimized | DEFault