Command Reference
R&S
®
ZNA
1247
User Manual 1178.6462.02 ─ 12
LTSTamp
Timestamp in local time
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
Starts the source calibration
●
at physical port <PhyPt>
●
in channel <Ch>
●
at harmonic orders
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
Then stores and applies the calibration data.
Use
SOURce:POWer:CORRection:PMETer:ID
to select the suitable power meter.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Port number
Example:
*RST; FREQ:STAR 0.5E+9; STOP 1E+9
Reset the analyzer (activating a frequency sweep) and set the
sweep range between 1 GHz and 1.1 GHz.
CALCulate:PARameter:MEASure '<TraceName>',
'h3b2(p1)'
Configure a 3rd harmonic measurement with source port 1 and
receive port 2.
SOUR:POW:CORR:PMET:ID 1
Select power meter no. 1 (previously configured in the External
Power Meters dialog and properly connected) for the RF source
calibration.
SOUR:POW1:CORR:HARM
Start source power calibration at source port 1.
CORR:POW:HARM:ACQ
Start receiver power calibration at receive port 2.
Usage:
Event
Options:
R&S
ZNA-K4
Manual operation:
See
"Harmonic Orders to Calibrate"
SOURce<Ch>:POWer:CORRection:IMODulation:PORT
<SourceForRecCal>
Selects the source port for the receiver power calibration of an intermodulation mea-
surement. An external generator must be configured before the generator port is avail-
able (
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
SCPI Command Reference