Command Reference
R&S
®
ZNA
1110
User Manual 1178.6462.02 ─ 12
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
Completes a fixture compensation, storing and applying the acquired data.
Suffix:
<Ch>
.
Channel number of the corrected channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
Prepares the analyzer for fixture compensation comprising a single or a series of fix-
ture compensation sweeps
(
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
). Previous compen-
sation data is deleted.
Suffix:
<Ch>
.
Channel number of the corrected channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected
<CalGroupFile>,
<Standard>, <TestPort1>[, <SecondPortOrAdapter>]
Reloads a set of previously acquired calibration data for a particular standard from a
file in the cal pool. The loaded data may be combined with new calibration measure-
ment data (
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
) in order
to simplify and speed up the new calibration procedure. The channel settings for loa-
ded and new calibrations (e.g. the number of sweep points) must be identical.
Note:
●
The analyzer performs a consistency check for the loaded data. If the loaded file is
incompatible with the channel settings of channel
<Ch>
, or if it does not contain
data for the specified standard and port(s), a command error message (
–
100,
"Command error;...") is generated.
●
For a sliding match, the R&S
ZNA can acquire and load measurement data for up
to 20 positions per port. Multiple calls of
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected <CalGroupFile>,
SLIDe, <TestPort>
for the same cal group file and test port, implicitly increases