Command Reference
R&S
®
ZNL/ZNLE
934
User Manual 1178.5966.02 ─ 07
POWer:S<Ptout><Ptin>
S-parameter with output and input port number of
the DUT, e.g. S11, S
21
.
POWer:RATio A<Ptout> | B<Ptin>,
A<Ptout> | B<Ptin>
Output:
A<Ptout>/B<Ptin>
Ratio, e.g. B2, A1 for b
2
/a
1
drive Port 1
POWer:A<Ptout>
Wave quantity with stimulus port number of the ana-
lyzer, e.g. a
1
.
POWer:B<Ptin>
Wave quantity with receive port number of the ana-
lyzer, e.g. b
2
.
POWer:Z<Ptout><Ptin>
Matched-circuit impedance (converted Z-parameter)
with output and input port number of the DUT, e.g.
Z
11
, Z
21
.
POWer:Y<Ptout><Ptin>
Matched-circuit admittance (converted Y-parameter)
with output and input port number of the DUT, e.g.
Y
11
, Y
21
.
POWer:KFACtor
POWer:MUFactor<Lev>
Stability factor K
Stability factors
μ
1
or
μ
2
VOLTage[:DC] DCIN1 | DCIN2
Output:
DC 1 V, DC 10 V
DC Input 1 or 2
Note:
The mnemonics POWer: and VOLTage: are not used in output strings.
TRACe:CLEar
<MemTrace>
Deletes one of the memory traces
Mem<n>[Trc<m>]
, where n = 1, ... 8.
Setting parameters:
<MemTrace>
MDATa1 | MDATa2 | MDATa3 | MDATa4 | MDATa5 | MDATa6 |
MDATa7 | MDATa8
Identifier for the memory trace; see
.
Range:
MDATA<n> where <n> = 1 to 8.
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1
).
TRAC:COPY "Mem_Pt20",CH1DATA
Copy the current state of the created trace to a memory trace
named "Mem_Pt20". The memory trace is not displayed.
DISP:WIND:TRAC2:FEED "MEM_PT20"
Display the created memory trace in the active diagram area
(diagram area no. 1).
Usage:
Setting only
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH