![Rohde & Schwarz 1323.0012.03 User Manual Download Page 657](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777657.webp)
Command Reference
R&S
®
ZNL/ZNLE
657
User Manual 1178.5966.02 ─ 07
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3, 1, 2000000000, 3000000000, 5
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
Define and enable a second ripple limit range in the stimulus
range between 2 GHz and 3 GHz, assigning a ripple limit of +5
dB.
CALC:RIPP:DISP ON
Show the ripple limits in the active diagram.
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DELete:ALL
Deletes all ripple limit ranges.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,
3, 1, 2000000000, 3000000000, 5
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
Define and enable a second ripple limit range in the stimulus
range between 2 GHz and 3 GHz, assigning a ripple limit of +5
dB.
CALC:RIPP:DEL:ALL
Delete both created ripple limit ranges.
Usage:
Event
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DISPlay[:STATe]
<Boolean>
Displays or hides all ripple limit lines (including all ranges) associated to the active
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - ripple limit line on or off.
*RST:
OFF
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:DISP ON
Show the ripple limit range in the active diagram.
Manual operation:
See
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH