Getting Started 1177.5756.02 ─ 13
Risk of damage to the R&S
FSWP
The maximum supported reverse power is 30
dBm. Make sure that this
value is not exceeded to avoid damage to the R&S
FSWP.
The maximum DC output voltage is 0
V.
For details on configuring the signal source output and generating a signal, see
the User Manual.
4.1.9
USB Ports
The three USB ports on the front panel (type A) allow you to connect devices like
keyboards, mouses or memory sticks.
The R&S
FSWP provides additional USB ports (including one type B port)
on the rear panel.
All USB ports support standard 2.0.
4.1.10
Power Sensors
The female Lemosa connector allows you to connect power sensors of the
R&S
NRP-Zxy family to the R&S
FSWP.
For an extensive list of supported power sensors, refer to the data sheet.
For details on configuring and using power sensors, see the User Manual.
4.1.11
Probe Power Connector (3 Pins)
The Probe Power connector allows you to supply external equipment with power
(for example probes or transducers).
The three-pinned probe power connector supports supply voltages from +15
V to
-12.6
V and ground. The maximum permissible current is 150
mA. This probe
power connector is suitable, for example, for high-impedance probes.