Remote Commands
R&S
®
Spectrum Rider FPH
501
User Manual 1321.1011.02 ─ 11
Parameters:
<DecimalValue>
Range:
0 to 65535
Example:
STAT:QUES:NTR 65535
STATus:QUEStionable:POWer:NTRansition
<DecimalValue>
This command controls the Negative TRansition part of the QUEStionable register.
Setting a bit causes a 1 to 0 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Parameters:
<DecimalValue>
Range:
0 to 65535
Example:
STAT:QUES:NTR 65535
STATus:QUEStionable:PTRansition
<DecimalValue>
This command control the Positive TRansition part of the QUEStionable register.
Setting a bit causes a 0 to 1 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Parameters:
<DecimalValue>
Range:
0 to 65535
STATus:QUEStionable:FREQuency:PTRansition
<DecimalValue>
This command control the Positive TRansition part of the QUEStionable register.
Setting a bit causes a 0 to 1 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Parameters:
<DecimalValue>
Range:
0 to 65535
STATus:QUEStionable:LIMit:PTRansition
<DecimalValue>
This command control the Positive TRansition part of the QUEStionable register.
Setting a bit causes a 0 to 1 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Parameters:
<DecimalValue>
Range:
0 to 65535
Remote Control - Commands